| GC-906 | |||
| Host | Pressure/Flow Control | Supports up to 18 channels of electronic pressure/flow control (EPC) | |
| Pressure Control Accuracy | 0.001psi | ||
| Number of sample ports supported | 3 | ||
| Maximum number of detectors supported | 4 | ||
| Temperature Control System | 8-channel independent temperature control system | ||
| Smart Maintenance Tracking and Reminders | yes | ||
| Column oven | Control temperature | Room temperature+2℃ ~ 450℃ | |
| Minimum step value | 0.01℃ | ||
| Maximum heating rate | ≥250ºC/min | ||
| Cooling time | From 450°C to 50°C in <3.2 minutes | ||
| Programmed Temperature Ramp | ≥99 steps 100 Platform | ||
| Temperature Stability | Ambient temperature ±1°C, column oven temperature variation <0.01°C | ||
| Others | 1. Automatic furnace chamber temperature protection function 2. Dual column box independent temperature control |
||
| Injection System | Injection type | Compatible with split/splitless capillary injectors, septum purging packed column injectors, cold column head injectors, programmable temperature vaporization injection ports, valve injection systems, headspace injection systems, thermal desorption injection systems, etc. | |
| Maximum operating temperature | 450℃ | ||
| Programmatic current increase | ≥ 10 steps | ||
| Stem pressure | 0~50psi、0-200psi | ||
| Pressure Control Accuracy | 0.001Psi | ||
| Maximum bypass ratio | 1:10000 | ||
| Flow Setting Range | 0~500ml/min 0~1250ml/min |
||
| Flow Setting Accuracy | 0.01ml/min | ||
| Carrier Gas Control Mode | Constant pressure, constant flow rate, constant linear velocity, programmed pressure, programmed flow rate, programmed linear velocity, pulse injection | ||
| Others | Features an independent split cold trap and a septum purge trap | ||
| Detector | Hydrogen Flame Ionization Detector (FID) | Maximum operating temperature | 450℃ |
| Detection limit | ≤1.1pg C/s(n-hexadecane) | ||
| Baseline drift (30 min) | ≤2×10⁻13A | ||
| Baseline noise | ≤2×10⁻14A | ||
| Dynamic Linear Range | ≥10⁷ | ||
| Function | Auto-ignition function | ||
| Thermal Conductivity Detector (TCD) | Sensitivity | ≥12000mV.mL/mg | |
| Baseline drift (30 min) | ≤0.2mV | ||
| Baseline noise | ≤0.05mV | ||
| Dynamic Linear Range | ≥105 | ||
| Function | Air Gap Flow Protection Function | ||
| Flame Photometric Detector(FPD) | Maximum operating temperature | 450℃ | |
| Detection limit | P≤80fg/s(Methyl parathion)、S≤8pg/s (Methyl parathion) | ||
| Dynamic Linear Range | P≥103;S≥102 | ||
| Function | Auto-ignition function | ||
| Electron Capture Detector (ECD) | radioactive source | 63Ni | |
| Maximum operating temperature | 400℃ | ||
| Detection limit | ≤1.0×10-14 g/mL(γ-666) | ||
| Dynamic Linear Range | ≥104 | ||
| Nitrogen-Phosphorus Detector (NPD) | Maximum operating temperature | 400℃ | |
| Detection limit | N:≤1×10-12g(N)/s(Azonaphthalene) P:≤5×10-13g(P)/s(Malathion) | ||
| Dynamic Linear Range | N≥103、P≥103 | ||
The GC-906 is a gas chromatograph designed for food testing, agricultural analysis and quality-control laboratories in Riyadh, Dammam, Dhahran, Jubail, Al Khobar, Jeddah, Makkah, Yanbu and Jazan across Saudi Arabia.