Dry Bath Incubator is designed for outstanding temperature control and accuracy. Digital wheel enables accurate setting of temperature and heating time. The equipment comes with a built-in temperature sensor and also an additional external temperature sensor PT1000. When PT1000 is connected, the internal temperature sensor will disconnect automatically, enabling the temperature control and display to rely on the external temperature sensor. Interchangeable aluminum alloy blocks provide versatility in applications.
| Model | TC0401003 |
|---|---|
| Number of blocks | 4 |
| Power [W] | 430 |
| Heating temperature range [°C] | room temp.+5°C-150°C |
| Temperature display | LED |
| Adjustment and display resolution [K] | 0.1 |
| Connection for ext. temperature sensor | DIN 12 878 |
| PT 1000 variation; DIN EN 60751 Kl. A [K] | < = ± (0.15 + 0.002 x ITI) |
| Temperature stability within the blocks at 37°C | ±0.5°C |
| Temperature stability within the blocks at 60°C | ±0.5°C |
| Temperature uniformity at 37°C | ±0.5°C |
| Heating rate / Heat up time withexternal sensor * [K/min] | 4 |
| Set-up plate material | Aluminium alloy |
| Set-up plate dimensions [mm] | 96×304 |
| Fixed safety circuit [°C] | 170 |
| Timer | √ |
| Time setting range | 1min – 99h59min |
| Dimensions (W x H x D) [mm] | 152 x 86 x 465 |
| Weight [kg] | 7.5 |
| Permissible ambient temperature [°C] | 5 – 40 |
| Permissible relative humidity | 80% |
| Protection class according to DIN EN 60529 | IP 21 |
| Voltage [V] | 100-120VAC/200-240VAC;50Hz/60Hz |
| Frequency [Hz] | 50/60 |
Supplied to order to laboratories across Saudi Arabia. Contact AL MUSK Union Scientific for specifications and a quotation.
Dry Bath Incubator is designed for outstanding temperature control and accuracy. Digital wheel enables accurate setting of temperature and heating time
0.00 ر.س