Technical Specifications

Specification Details
Model MS-H280-S6
Cat.No. 8030351112
Work plate Dimension Ø134mm
Work plate material Aluminum with ceramic coating
Max. stirring quantity [H2O] 3LX6
Speed range 200-1500rpm
Speed display LCD
Temperature display LCD
Heating temperature range 25-280℃
Temperature control accuracy ±1℃(<100℃), ±1%(>100℃)
Heating temperature control accuracy test conditions 1 L beaker with 500 mL water, 50 °C, 40 mm magnetic stir bar, 500 rpm; 1 L beaker with 500 mL silicone oil, 120 °C, 40 mm magnetic stir bar, 500 rpm.
Temperature display accuracy 0.1℃
External temperature sensor PT1000(accuracy±0.2℃)
“Hot” warning 50℃
Overheating temperature protection 420℃
Motor type Brushless DC motor
Stirring positions 6
Max. magnetic bar length 40mm
Protection class IP21
Motor rating output 15WX2
Heating power 1800WX2
Power 1815WX2
Voltage,Frequency 200-240V,50Hz
Dimension [WxDxH] 564X523X127mm
Weight 12.7Kg
Permissible ambient temperature 5-40℃
Permissible ambient humidity 80% RH

Description

The 4/6 Channel LCD Digital Hotplate Magnetic (MS-H280-S6) is a magnetic hotplate stirrer supplied for laboratory and research use in Riyadh, Dammam, Dhahran, Jubail, Al Khobar, Jeddah, Makkah, Yanbu and Jazan across Saudi Arabia.

  • Product Features:
  • 4/6 channels with independently control of start/stop, temperature, speed and display for better parallel experiment control.
  • Ceramic coating aluminum work plate with fast heat transmission and strong resistance to corrosion.
  • Powerful, rapid heating with accurate PID temperature control
  • Spark-free DC brushless motor with extended life and no maintenance
  • Compatible with various heating block or reaction vessels for a variety of application scenario
  • Ultra space-saving with 4/6 channel integrated in one instrument
  • “HOT” warning flashes on the LCD display if the work plate temperature is above 50℃ after the user turn off the power switch.
  • Two independent safety circuits keep monitoring the instrument.When overheating is detected and the sensor malfunctions, it will stop heating and sound alarm; When short circuit is detected, it will cut off the power to
  • Reliable safety mechanism, compliant to IEC61010-1UL61010-1BSEN61010-1CANCSA-C22.2NO.61010-1and all other safety regulations of laboratory heating devices, competent with tasks without human surveillance.